Hematology
Equipment and Resources
HEM-ERHaematology runs on instruments that the people at the bench must start up, calibrate, maintain and rescue when they misbehave. This domain covers routine operation and preventive maintenance of cell counters, coagulation analysers, centrifuges and microscopes, together with the structured troubleshooting that separates an instrument fault from a genuine patient result. It also covers the capture of digital images that are clear enough, and identified well enough, to stand as part of the laboratory record. Equipment and Resources is one of the eight competency areas assessed in the Hematology Fields-of-Practice examination set by CAMLPR.
Running, calibrating and maintaining haematology instruments
Cell counters, coagulation analysers, centrifuges and microscopes only produce trustworthy numbers when they are set up and looked after properly. A shift begins with start-up checks and background counts, and the technologist confirms that the analyser's calibration is current against the manufacturer's schedule. Recalibration follows a major service, a change of calibrator lot, or a control shift that nothing else explains. Daily, weekly and monthly maintenance — probe cleaning, aperture flushing, bath and filter changes, rotor inspection — is performed and signed off. Every calibration, service call and maintenance task is recorded, because an audit asks for evidence rather than recollection.
A field service engineer has just replaced the optical flow cell in a haematology analyser. Both levels of quality control recover within range on the first run afterwards, and the technologist has a bench full of waiting samples. What should be done before patient results are released?
At morning start-up an impedance cell counter reports a white cell background count of 0.4 ×10⁹/L, above the manufacturer's stated ceiling of 0.2 ×10⁹/L. Reagent containers are full and in date, and the diluent lot was opened yesterday. Which action addresses the most likely cause?
Spotting instrument faults and working through them methodically
Instruments fail in ways that look like patient results, so the first skill is noticing that a value, a flag or a control shift does not fit. From there the technologist follows a structured path: read the flag text and error log, inspect the specimen for clots or short fill, repeat the test, check reagent lots and volumes, look for bubbles, worn tubing or an obstructed aperture, and run the manufacturer's diagnostic routines. Anything unresolved is escalated to service, testing is diverted to a backup analyser so turnaround is protected, and the fault, the steps taken and the outcome are documented.
After a leaking wash block on a cell counter was replaced, the supervisor wants proof that sample is no longer being carried from one aspiration into the next. Which procedure demonstrates that carryover is back within the manufacturer's acceptable limit?
The chart shows a linearity study on an analyser's platelet channel: recovered values track the assigned values closely up to roughly 800 ×10⁹/L and then flatten out. A specimen from a patient with a myeloproliferative disorder returns a platelet count of 940 ×10⁹/L with no instrument flag. What is the appropriate response?
Capturing digital images of films and marrow preparations
Haematology increasingly photographs findings rather than describing them: digital cell imagers locate and pre-classify leucocytes, and cameras record blood films, marrow aspirates and stained preparations for consultation, teaching and the permanent record. The technologist sets the optics so that morphology and colour are reproduced faithfully — correct objective, sharp focus, properly adjusted condenser, white balance set on a blank field — and avoids heavy compression that smears granularity and chromatin detail. The field chosen must be representative rather than flattering. Each captured frame is linked to its accession identifier and date, stored where it can be retrieved, and treated as patient information.
Images are being captured from a Wright-stained film so a haematologist at another site can review the neutrophil granularity. On the monitor the first captures look distinctly bluish and the cell outlines are soft. Which adjustment should be made before the frames are saved?
A set of marrow aspirate images captured during a morning run is being filed so that it can be pulled up months later if the case is reviewed. Which practice keeps the image set usable as part of the permanent record?
Back to all hematology areas in the Hematology study guide.
MLTPrep is an independent study resource. Not affiliated with or endorsed by CAMLPR. Always confirm current requirements with CAMLPR directly.